Technical data
13.2 Electrical, EMC and climatic requirements
Electromagnetic compatibility
Table 13- 13 Requirements: Interference emission in industrial area in accordance with EN 61000-6-4
Comments
Emission of radio interferences
(electromagnetic fields)
Emission on power supply cables
24 V
Emission conducted Ethernet
170
Standard
Class A industrial environment:
EN 61000-6-4
IEC/CISPR 16-2-3: 2008
Class A: Industrial environment:
EN 61000-6-4
IEC/CISPR 16-2-1: 2010;
EN 55016-2-1: 2009
EN 61000-6-4
Limits
30 ... 230 MHz,
•
40 dB (μV/m) Q
230 ... 1 000 MHz,
•
47 dB (μV/m) Q
Class A: Industrial environment
0.15 ... 0.5 MHz, 79 dB (μV)
•
Q
0.15 ... 0.5 MHz, 66 dB (μV)
•
M
0,5 ... 30 MHz, 73 dB (μV) Q
•
0,5 ... 30 MHz, 60 dB (μV) M
•
0.15 ... 0.5 MHz:
53 dB (μA) to 43 dB (μA) Q
•
40 dB (μA) to 30 dB (μA) M
•
0.5 ... 30 MHz:
43 dB (μA) to 30 dB (μA) M
•
SIWAREX WP231