GE Sievers InnovOx Operation And Maintenance Manual page 181

On-line toc analyzer
Hide thumbs Also See for Sievers InnovOx:
Table of Contents

Advertisement

Chapter 9: Troubleshooting
6436 – Reactor Leak Error
Message: "Go to the Diagnostics screen and run the Gas Leak tests. If any of the tests fail, please export the
errors and contact GE Analytical Instruments for further instructions."
Explanation: A leak in the Reactor has been detected. The fittings on the ends of the Reactor tube may be
leaking, or there may be a leak in the Reactor tube itself.
Other software errors are listed in Table 13, "Other Software Errors," on page 182. These errors are grouped
according to the module in which the error is found.
6437 – Invalid Conversion Equation Error
Message: "Conversion factor resulted in a mathematical error while applying the equation."
Explanation: When the equation assessment with your TOC measurement is invalid, this message will appear.
Please review your equation.
6438 – Printer Error
Message: "System either failed to detect an HP printer, or the printer is not currently supported. Please export
the errors and diagnostics, and contact GE Analytical Instruments for further instructions."
Explanation: Please refer to Chapter 7 for a list of supported printers.
6440 – Datastore Manually Archived
Message: "Datastore manually archived by user. Historical marker only, no follow-up necessary."
Explanation: This warning is meant as an historical marker for when a user manually archives the datastore.
GE Analytical Instruments ©2013
181 of 238
DLM 68100-06 EN Rev. A