Performance Tests
4-1 2. PERFORMANCE TESTS
T h e Performance Tests are divided into 6 main tests, D C Volts, AC Volts, D C Current, AC Current, Ohms,
and Frequency. Each test can be performed independently of the others. T h e steps within each test,
however, should be performed in order.
- -
Figure 4-1. DCV Performance Test Connections
I
NOTE
I
The temperafurr o f fhc cnvrronmrnf whtw t h o r fests are f o he pc~rformt~d ml/.u he'
wilhin +5
C of the /cwpcratl/rc whcrc f h t ~ in.s/rlimcw/ war culihra/c~d. Thc incfrurncwt
was calihrated at the factory in an area with a fempcratl/rr o f
24
O
C
( f
I
O
C).
4-1
3.
Preliminary Steps
1. Turn the instrument
ON
and allow a o n e hour warm-up period.
2.
Be certain all external inputs are disconnected f r o m the H P 3457A. Run the
TEST
routine. T h e test
result must read
"SELF TEST OK".
If the display shows
"TEST FAILED",
the instrument probably needs
repair.
3. Run the
AUTO CAL
routine. (Press the
AUTO CAL
key, enter the number 1 and press the
ENT
key).
DC Voltage Performance Tests
4-14. DC Voltage Function
-
Offset Test
4-1 5. Equipment Required. A low thermal short (copper wire) is required f o r this procedure.
I.
Connect a low thermal short across the Front Panel
HI
and
LO
Input Terminals.
2.
Set the
HP 3457A t o the DC Voltage function
(DCV)
and the number of digits displayed
(DIGITS
DISP)
t o six.