Process validation
The standard of disinfection in the
disinfection programmes must be
confirmed by the user as a routine
matter.
In the UK, safety checks should be
carried out to the machine every 3
months by a Miele approved service
technician in accordance with EN ISO
15883 and HTM 2030.
Routine checks
Before the start of each working day,
the user must carry out a number of
routine checks. A check list is supplied
with the machine for this purpose.
The following need to be inspected:
– All filters in the wash cabinet
– The spray arms in the machine and
in the baskets
– The wash cabinet and the door seal
– The dispensing systems
– Baskets and inserts
Maintenance
43