Transceiver Installation Requirements For Multiple Itss; Patient-Worn Transceivers; Extra Mac Addresses In The Label Assignment Screen - Philips IntelliVue Telemetry System Infrastructure Installation And Service Manual

Intellivue telemetry system infrastructure
Table of Contents

Advertisement

Transceiver Installation Requirements for Multiple ITSs

New installations or replacement of 1.4/2.4 GHz patient-worn transceivers require
an association process to establish connection with the ITS infrastructure and the
ICN Central Station to get an equipment label assigned. Normal monitoring cannot
proceed without the transceivers being assigned an equipment label.

Patient-worn Transceivers

When transceivers are shipped from the factory, they are shipped with an
Equipment Label of "New Device" and an RF Access Code of "0". This allows them
to connect to any access point (AP). However, a new transceiver will connect to the
first access point it detects—not necessarily the AP that is closest.
If you do not see the MAC address appear in the Label Assignment screen (), then it
is possible that the transceiver is connected to an access point associated with a
different network. You will need to reboot the transceiver and then click the
"Refresh" key until you see the MAC address appear in the Label Assignment
Screen.

Extra MAC addresses in the Label Assignment Screen

if the systems are so close that it is likely that transceivers will connect to another
system during the installation process, then it is likely that you will see extra MAC
addresses in the Label Assignment screen. These would be transceivers from the
other system that connected, but were not assigned a label because they
connected to the wrong system.
If a user wants to move an assigned portable device to a new area with a different
RF Access Code, then the portable device must be re-configured as a 'new device'
with no equipment label assigned, And, at this stage the normal process to assign
the portable device to the new AP group and Clinical care unit can take place as
defined above.
IntelliVue Telemetry System Infrastructure Installation and Service Guide
Transceiver Installation Requirements for Multiple ITSs
Figure A-2: Assigning Labels to Patient-worn Transceivers
A-3

Hide quick links:

Advertisement

Table of Contents