Hitachi TM3000 Instruction Manual page 225

Tabletop microscope
Table of Contents

Advertisement

11. Attaching the cap Secure the cap, by turning it clockwise, onto the cartridge filament
that was attached in Section 5.1 10. .
Figure 5.1-20 Cap
12. Verify that grease is applied to the o-ring and is free of particles, then attach the o-ring
to the groove.
O-ring
NOTE : Thinly apply the supplied vacuum grease to the entire o-ring. Failure to properly
apply grease to the o-ring can cause the air, trapped in the o-ring groove, to enter
the electron gun when the electron gun is moved during the beam axis adjustment
process, resulting in the beam being turned off automatically.
Cap
Figure 5.1-22 Electron Gun O-ring
5 - 8
Figure 5.1-21 Attaching the Cap