Modei5065A
Circuit Diagrams, Theory, and Maintenance
SECTION Ill
OPERATION
3·1.
INTRODUCTION
Table 3-1 Operating Checks
3-2.
This section provides operating procedures for the
5065A Rubidium Vapor Frequency Standard. Tables 3-1
and 3-2 gives the basic turn-on procedure. Figures 3·9,
3·10, and 3·11 explain front, top, and rear controls and
connectors.
3·3.
OPTIONS
001
AND
002
(Option
003
=
001 and 002)
3·4.
Operating procedures lor Option 001 (Time Stand-
ard) and Option 002 (Standby Power Supply) are covered
in Paragraphs 3-19 through. 3-31.
3-5.
OPERATING PROCEDURE
3-6.
General
3-7.
In instruments equipped with Option 002, Standby
Power Suply, remember that the internal standby battery
is fully discharged when delivered and must be brought
to full charge (16 hours minimum) before it can deliver
rated standby power. Battery charging instructions are
included in Figure 3-2, Turn-On Procedure. For more
standby power, available accessories are the HP Model
5085A Standby Power Supply or HP Model K02-5060A
Power Supply.
3-8.
Turn-On Procedure (see Figure 3-2)
3-9. Turn-On After Long Storage
If the 5065A has been in storage for longer than 2 months,
there is a possibility of cell flooding occurring in the RVFR
tube. If after 1 hour of warm-up from initial
tum-on
no
2nd harmonic is present, then cell flooding can be sus-
pected. The following procedure should be used to incor-
rect cell flooding.
a.
Remove all power from 5065A and remove bot-
tom cover.
b.
Disconnect the red and black twisted pair of wires
from XA4( 1
1
and XA81 11 respectively.
c.
Obtain a current-limiting power supply capable
of producing 1 amp • power supply voltage is not important).
d.
Before turning on power supply, place a short
across its output terminals.
e.
Connect the power supply as shown in Figure 3-1.
The short should remain in place on the power supply
output.
f.
Set power supply voltage and current controls to
minimum position. The precautions of stepsd, e, and fare
to prevent the filter capacitor on the power supply output
from discharging into the RVFR.
CIRCUIT CHECKS
Switch Position
Meter Indication
BATTERY
35-45
Option 002)
SUPPLY
38-42
LAMP OVEN
10-45
CELL OVEN
10-45
OSCOVEN
25-50
PHOTO I
25- 50
5 MHz
35-45
(no load)
CONTROL
-50 to +50
ERROR
0
2ND HARMONIC
20-50
1 MHZ
38- 42
(no
load)
100kHz
38-42
(no
load)
Description
Indicates
battery voltage
Indicates +20
volts regulated
supply
Indicates
power to lamp
oven in RVFR
Indicates
power to
absorption
cell oven in
RVFR
Indicates
power to
quartzose
oven
Indicates
RVFR output
current
Indicates
5 MHzoutput
level
Indicates de
control volt-
age !.2-9uartz
oscillator
Indicates fre-
quency dif-
terence
between RVFR
and micro-
wave field as
a de voltage
Indicates 2ND
HARMONIC
level
Indicates 1
MHz output
level
indicates 100
kHz output
level
3-1