Cleaning, Maintenance, Disposal
Calibration Mode
(e.g. to check the pneumatic system for leaks)
 Connect a rubber bulb between pressure tubing and
cuff, using a T-adapter.
 Roll up cuff tight.
 Switch off device and switch it on again after a few
seconds.
 Wait for time to be displayed.
 Push
 Push
must be between 25 and 100. If the displayed value is
outside this range, TONOPORT V must be returned
for repair.
 Push
play now indicates the pressure in mmHg).
 Generate a test pressure of 200 mmHg and measure
the pressure decrease after waiting at least 30 sec-
onds. (Pressure decreases between 3 and 5 mmHg are
typical; a pressure decrease > 6 mmHg indicates a
leak and the system needs to be repaired.)
 Push
View Firmware Version
 Turn on the device and wait for the time to be dis-
played.
 Push
 Push
– "21:00" =
 Push
TONOPORT V
INFO
START
four times: the display indicates "H 4".
STOP
INFO
: the display indicates an internal value that
START
START
STOP
STOP
INFO
again: the display indicates "0" (the dis-
START
START
STOP
STOP
INFO
START
START
to exit the calibration mode.
STOP
STOP
INFO
five times: the display indicates "H 5".
START
STOP
INFO
START
START
: the firmware version is indicated, e.g.
STOP
STOP
firmware version 2.1
INFO
to end the display.
START
START
STOP
STOP
2001589-085 Revision G