USB Port Verification
For verification of the USB port, turn on TONOPORT V
and connect its USB port to the PC.
1.
Start the Device Manager of the operating system.
2.
Double-click Ports (COM and LPT) to view all
ports.
If CardioSoft version 6.72 or higher is used or if a
USB serial port (TUSB3410 Device) between COM1
and COM4 has been selected, it is not necessary to
choose another port. Note down the selection, because
the same port must be set in CardioSoft, and close all
windows to return to the Windows desktop.
If a USB serial port greater than COM4 has been
selected one of the ports COM1 through COM4 must be
disabled at the PC so that it can be assigned to the USB
port.
3.
Select one of the ports COM1 through COM4 that
is not needed for other devices and disable it (right-
click > Disable). Confirm the message that the
device will not be functional any longer.
4
Right-click USB - Serial Port (COM X) and click
Properties.
5.
Click Port Settings > Advanced and at COM Port
Number select the port that you disabled earlier.
Select Yes, if a message appears that this COM port
is already used by another device. The selected port
must also be set in CardioSoft.
6.
Select OK where needed and/or close all windows
to save the settings.
7.
Disconnect the USB cable and restart Windows.