Sweep Oscillator.
RF Plug-In...
Source Module.
Directional Coupler (two required)
Detector (two required plus DUT)
Scalar Network Analyzer.
Short.
Low Pass Filter.
Isolator.
. .HP8350B
.HP 83550A
. HP 83554A
. . HP R752C
HP R85026A
. . . HP 8757A
..HPR921A
. .HP R362A
. .HP R365A
1.
Connect equipment as shown in Figure 3-1, with the short connected to the coupler, and allow the
equipment to warm up for 30 minutes.
On the HP 8757A:
2.
Press [PRESET] to preset both the HP 8757A and HP 8350B.
On the HP 8350B:
3.
Press:
[START] [2] [6] [.] [5] [GHz]
[STOP] [4] [0] [GHz]
This step sets the frequency range at the output of the HP 83554A source module.
On the HP 8757A:
4.
Select
[CHAN 2 OFF],
5.
Press [MEAS] and select [A/R] softkey.
6.
Press [DISPLAY] and select [MEAS-*-MEM] to store the channel 1 display in memory.
7.
Select [MEAS-MEM]. A flat trace Is displayed on the CRT.
8.
Connect the detector to be tested in place of the short in the test setup.
On the HP 8757A;
9.
Press [SCALE] and select [AUTOSCALE] in the SCALE menu.
10. Press [CURSOR] and select the [MAX] softkey.
11. Read the worst case return loss from the cursor value given in the active entry area of the CRT.
Record this value on the test record card.