Instrument Test
A self-test may be started by pressing and holding
both the LAuto) and @G&i) keys while turning on the
cardiograph with the (On/Standby) key. The instrument
performs an internal self-test. The test results are
printed on the report for use by service personnel. This
self-test runs continuously until the cardiograph is
turned to Standby (off).
Patient Safety
Patient Isolation: Less than 20 PA leakage with 120
VAC, 60 Hz or less than 50 PA leakage with 240 VAC,
50 Hz with patient cable.
Defibrillation Protection: Protected against damage from
400 joule defibrillator discharges.
Power and
Environment
Line Power: 100 to 120 VAC, 50 to 60 Hz (at 115 VAC
power-switch setting); 220 to 240 VAC, 50 to 60 Hz (at
230 VAC power-switch setting); 50 VA Maximum.
Fuse: 0.4 A for both the 115 and 230 VAC switch
settings.
Environmental Operating Conditions: 10 to 40" C (50 to
104" F), 15 to 80% relative humidity, non-condensing, up
to 4,550 m (15,000 ft.) altitude.
Environmental Storage Conditions: 0 to 50" C (32 to
122" F), 15 to 90% relative humidity, non-condensing, up
to 4,550 m (15,000 ft.) altitude.
Cardiograph Dimensions: 43 by 39 by 10 cm (17.0 by
15.3 by 3.9 in.).
Cardiograph Weight: 8.5 kg (19.8 lbs.).
B-2 Specifications