Care And Cleaning; Cleaning The Cardiograph - HP PageWriter 100 User Manual

Table of Contents

Advertisement

Maintaining the Cardiograph
Care and Cleaning
The outside surfaces of the cardiograph and its
accessories (except the patient cable) are designed
to be cleaned by mild soap and water or isopropyl
alcohol. The patient cable can be cleaned only with mild
disinfectant OT soap and water. The patient cable cannot
be cleaned with isopropyl alcohol.
Cleaning the
1. Unplug the power cord and ensure that the
Cardiograph
cardiograph is in Standby mode (all lights are off).
2. Wipe the external surfaces of the cardiograph with a
soft cloth dampened with mild soap and water or
isopropyl alcohol. Avoid applying cleaning fluids to
the cable connectors.
Caution
a
Do not use any strong solvents or abrasive cleaning
materials.
Do not spill any liquids on the surface of the
cardiograph. Immediately have the cardiograph
serviced if any liquids spill on the surface of the
cardiograph.
Do not use the following to clean the cardiograph:
n
Acetone
H Iodine-based cleaners
H Phenol-based cleaners
n
Ethylene Oxide Sterilization
n
Chlorine bleach
H Ammonia-based cleaners
Maintaining the Cardiograph 4-1