Siemens SITRANS LG260 Operating Instructions Manual page 33

Guided wave radar
Hide thumbs Also See for SITRANS LG260:
Table of Contents

Advertisement

Date/Time
Reset
PBD-51041061
Parameter settings for recording the echo curve and the settings of
the echo curve itself can be carried out in the adjustment software
PACTware.
With the adjustment software PACTware and the PC the high-reso-
lution echo curve can be displayed and used later on to assess the
quality of the measurement.
6.4.4 Additional adjustments
In this menu item, the internal clock of the sensor is set.
After a reset, certain parameter adjustments made by the user are
reset.
Note:
After this menu window, the reset process is carried out. No further
safety inquiry follows.
The following reset functions are available:
Delivery status: Restores the parameter settings at the time of ship-
ment from the factory, incl. order-specific settings. Any stored false
signal suppression or user-programmed linearisation curve, as well
as the measured value memory, are deleted.
Basic settings: Resetting of the parameter settings incl. special
parameters to the default values (presettings) of the respective instru-
ment. Any created false signal suppression or user-programmable lin-
earization curve as well as the measured value memory are deleted.
The following table shows the default values of the instrument. De-
pending on the instrument version or application, all menu items may
not be available or some may be differently assigned:
SITRANS LG260 - Operating Instructions
33