(2) Pipe Orientation
The detector may be installed in horizontal, vertical or sloping pipe runs as shown in
Figure 4.2. However, except for horizontal installation, fluid should flow from lower
to upper directions. If no air bubble, Vertical down flow application are acceptable
under pressured piping conditions. See Figure 4.2.
.
Detector
(a)
The electrodes should be positioned horizontally relative to the ground surface in any
piping installation. See Figure 4.3.
Detector
Figure 4.3 Installation position of the detector
(c)
Figure 4.2 Detector Piping Orientation
A
A'
Flow direction
(b)
(a) Horizontal pipe installation
(b) Vertical pipe installation
(c) Sloping pipe installation
Electrodes
Cross-section A - A'
15
6F8A0916
Ground surface
Ground surface