Lithium battery
Fig. 48: Changing the lithium battery
1. Press the catch in the direction of the arrow (1).
2. Remove the battery (2).
3. Push the new lithium battery of the identical type into the holder and press it downward
until it engages (3).
Restoring BIOS settings (necessary after battery replacement)
1. If necessary, send any patient images in the system to the archive or burn them to a
CD/DVD.
2. Install the BIOS according to the specific system documentation.
3. At some systems the BIOS installation is only possible together with the complete SW
installation procedure.
Siemens
Medical Solutions
The battery jumps out of the holder slightly.
NOTE
When the battery is empty, the BIOS is set to default.
In addition to activating the system ON/OFF switch, the fol-
lowing may be necessary ->
- Turn OFF the PC via the power switch at the rear panel of
the PC.
- Turn ON the PC via the rear power switch and press the
push button on the front panel of the PC.
TD00-400.841.40.02.02
01.08
Page 49 of 52
CS SD SP/CO
49