High Flow Rate; Additional Replacement Parts In L-2200U - Hitachi L-2200 Instruction Manual

Optional syringe autosampler
Table of Contents

Advertisement

2.2.6
Replacement
Parts for Use at

High Flow Rate

2.2.7
Additional
Replacement
Parts in L-2200U
In an application where a considerably high flow rate is used
(e.g., a flow rate of 6 mL/mm or higher), replace the stainless
pipe connected between the injection port and the injection
valve port 3 (P/N 890-3451, 0.17 mm ID × 0.36 m) on the
L-2200 autosampler with the stainless pipe (P/N 890-3473, 0.5
mm ID × 0.36 m) furnished together with the optional syringe.
(See Figs. 1-1, 1-2, and 1-3.)
Stainless pipe
Fig. 2-9 Replacement Parts in High-Flow-Rate Application
To install an optional syringe onto the L-2200U autosampler,
remove the stainless pipe (P/N 890-3452, 0.1 mm ID × 0.36 m)
connected between the injection port and the injection valve
port 3 on the L-2200U autosampler, and instead of it, connect
the stainless pipe (P/N 890-3451, 0.17 mm ID × 0.36 m)
included in the accessory box accompanying the L-2200
autosampler. (See Figs. 1-1, 1-2, and 1-3.)
NOTICE: If the optional syringe is used without
accomplishing the above-mentioned replacement
with the stainless pipe (P/N 890-3451, 0.17 mm ID
× 0.36 m), it may result in performance
degradation.
2 - 9