HP 8340B Manual page 20

Hide thumbs Also See for 8340B:
Table of Contents

Advertisement

HP-IB Address Access
1.
Press:
[SHIFT] [LOCAL]
This
lets
you see the HP-IB
address
in
the
ENTRY DISPLAY.
2.
To change the HP-IB address (00
to 30):
a.
Rotate the front panel knob or use the keyboard entry.
b.
When
the desired
address
is
set, press
[Hz]. The HP-IB
address
is
changed, and the CHECK¬
SUM updated
(note that only the working data
is
updated).
3.
To
write protect the HP-IB address,
set
bit
5
of
cal
constant
#57
and lock out the
[SHIFT] [LOCAL]
function:
a.
Access the working data
cal
constant #57 as described above.
b.
Set
bit 5
by adding
32
to
the
HP-IB address. Enter
this value.
c.
Press
[Hz].
Automatic SYTM to YTO Tracking (Power Optimization)
1.
To
recalculate and store the
working data checksum,
and update the
SYTM
GAIN calibration
constants (#9 through 12 and 50 through 53) for peak RF output power, press:
[SHIFT] [PEAK]
This also automatically turns
off
the
CAL FAULT
indicator without performing an instrument
preset.
Working Data to Protected Data Transfer
1.
To recalculate
and store the working
data checksum,
and
store working data as protected data,
press:
[SHIFT]
[MHz]
[1] [4] [Hz]
[SHIFT] [kHz]
[5] [3]
[4] [9] [Hz]
This
is
the
only
way
to to update protected data.
If
this
data transfer
fails,
the
message EEPROM
FAILURE, CAL NOT RESTORED
is
displayed
in
the ENTRY DISPLAY.
Protected Data to Working Data Transfer
1.
To
recall protected data
and store
it in
the working data
memory
location,
press:
[SHIFT] [1Hz] [1] [4] [Hz]
[SHIFT] [kHz] [1] [9] [4] [6] [Hz]2.
After
writing
over
the
working data, press [SHIFT] [PEAK] to optimize
instrument operation.
A-14