Siemens Healthineers ADVIA Quick Reference Manual page 20

Autoslide system

Advertisement

Troubleshooting
Error Code
ASL - Printer error
ASL - Rack Error - Stop
ASL - Rack Handling error
ASL - Slide Carrier error
ASL - Smearing module not primed
ASL - Smearing module not ready
ASL - Smearing tape is loose
Severity
Possible Cause
Failure
Printer ribbon not advancing or jammed
Failure
General rack problem, could be low on racks
Failure
Hardware failure in rack handler track
Failure
Slide did not make it to the carrier in time, or loaded
incorrectly
Failure
Smearing module is not primed
Failure
Smearing module problem
Failure
Smearing film empty, broken or incorrectly
threaded
Corrective Action
Inspect printer ribbon - clear jam or replace ribbon
Check printer area for glass particles
Reset Smearer
Replenish racks in dispenser
Check for jammed rack or full output rack collection area
Reset Smearer
Check for jammed racks or debris
Reset Smearer
Clear slide and inspect area for glass particles or debris
Reset Smearer
Prime Smearer
Reset Smearer
Perform Daily Startup if necessary
Perform Reset Smearer
Replace smearing film
Clean smearing wedge
Note:
If the Autoslide is powered off, please wait 30
seconds before the power is switched back on.
This allows the ADVIA software to timeout
appropriately.