4. Option of EUP-O732T

In case of using RVS (Real-time Virtual Sonography), confirm that
type of the magnetic sensor. There are two types of the magnetic
sensors for EUP-O732T, EZU-RV2S and EZU-RV3S. EZU-RV2S and
EZU-RV3S are shown in Fig.3 and Fig.8. The uses of EUP-O732T with
wither of the magnetic sensors enables the user to perform RVS
(Real-time Virtual Sonography).

4.1 Magnetic Sensor (EZU-RV2S)

Magnetic sensor (EZU-RV2S) as shown in Fig.3 is the magnetic sensor
(EZU-RV2S) for EUP-O732T.
4.1.1
How to Attach the Magnetic Sensor
The procedure of attaching magnetic sensor is as follow.
1) Confirm that the magnetic sensor attachment and the Spacer for
EZU-RV2S are disinfected and sterilized.
2) Connect the probe to the ultrasound diagnostic scanner and
operate the scanner. Then adjust the image according to the
instructions given in the operation manual for the scanner.
3) To use RVS (Real-time Virtual Sonography), attach the magnetic
sensor as shown below in the out of operative field.
a) Attach the Spacer for EZU-RV2S to Magnetic sensor. Then insert
Magnetic sensor into the Magnetic sensor attachment with the
correct direction as shown in Fig.4.
Magnetic sensor
attachment
Magnetic sensor
(EZU-RV2S: Black)
Projected part
Spacer for EZU-RV2S
Fig.3 EZU-RV2S
Fig.4 How to attach the Magnetic sensor
Magnetic sensor
(black)
Projected part
-7-
Projected
part
Q1E-EP1234