3. Operation Procedure

1) Confirm that the probe, the Magnetic sensor attachment and the
Spacer for EZU-RV2S are cleaned, disinfected and sterilized.
The Magnetic sensor attachment and the Spacer for EZU-RV2S are
needed for Real-time Virtual Sonography (RVS). Regarding the
option for RVS, please refer to "4.Option of EUP-O732T".
2) It is recommended to use a disposable probe cover for preventing
a patient from infection and the probe cover should be allergy
free material to avoid allergic reaction.
3) Connect the probe to the ultrasound diagnostic scanner and
operate the scanner. Then adjust the image according to the
instructions given in the operation manual for the scanner.
4) The relationship between the direction of the probe and the
right-left orientation mark on image is shown in Fig. 2.
Cable
Scan head
Fig. 2 Relationship between direction of the
5) Use under sterile condition, protecting the probe by using
covers is recommend. Some Latex material may create
allergic reaction. Please use allergy free material
covers.
probe and Right-left orientation Mark
Probe
housing
H
Ⓗ: Right-left
Image
-5-
orientation
mark
Q1E-EP1234