Oxley Creek WWTP Water Reclamation
(Probe LU Manual) Vendor Manual
Appendix G: Asynchronous
Communications Data M ap
Directory
Slot
Index Value
1
0
1
2
3
1
4
5
6
7
8
Physical Block 1
This is the only physical block for the device.
Relative
Name
Index
0
7ML19985JB02
Description
Directory Header
1
1 - Directory ID
1
1 - Revision Number
3
3 - Number of entries
8
8 - Directory Entries
1
1 - First Composite List
3
3 - Num Composite Lists
List Directory
Entry Index No. of blocks
0x01040001
1
0x02050002
2
0x03070002
3
Directory
Slot Index No. of parameters
0x00100053
0
0x0064001C
0
0x014D008A
1
0x0110002F
1
0x0210002F
2
Access Structure
Block Object
r
SITRANS Probe LU (PROFIBUS PA) – INSTRUCTION MANUAL
4
1
(1 PB, at Index 4)
5
2
(2 TBs, at Index 5)
7
2
(2 FBs, at Index 7)
16
83
Physical Block
100 30
(Internal Use Only)
77
138
Transducer Block
16
47
AIFB1
16
47
AIFB2
Reserved
Block Object
Parent Class
Class
DD Reference
DD Revision
Profile
Profile Revision
Execution Time
Number of Parameters
Index of View 1
Number of views
Doc Id: VM11
Active Date: 15/12/2006
Size
(bytes)
4
4
4
4
4
4
4
4
Type
UNSIGNED_FLOAT (2)
INTEGER (1)
INTEGER (1)
INTEGER (1)
UNSIGNED_INTEGER(4)
UNSIGNED_INTEGER(2)
INTEGER (2)
INTEGER (2)
UNSIGNED_INTEGER (1)
UNSIGNED_INTEGER(2)
UNSIGNED_INTEGER(2)
UNSIGNED_INTEGER(1)
Page 61