Important EMC notices for use in medical environments
• The HMI-3000MT needs special precautions regarding EMC and needs to be installed and put into service according to
the EMC information provided in the instructions for use.
• Portable and mobile RF communications equipment, such as cellular phones, can affect the HMI-3000MT.
WARNING
The use of accessories and cables other than those specified, with the exception of replacement parts sold by Sony
Corporation, may result in increased emissions or decreased immunity of the HMI-3000MT.
Type of cable
HMO-CA50M : Head Mounted Display Cable
Guidance and manufacturer's declaration – electromagnetic emissions
The HMI-3000MT is intended for use in the electromagnetic environment specified below. The customer or
the user of the HMI-3000MT should assure that it is used in such an environment.
Emission test
RF emissions
CISPR 11
RF emissions
CISPR 11
Harmonic
emissions
IEC 61000-3-2
Voltage
fluctuations/flicker
emissions
IEC 61000-3-3
WARNING
If the HMI-3000MT will be used adjacent to or stacked with other equipment, normal operation of the HMI-3000MT under
such configurations should be verified via observation.
4
List of cables used for EMC test
Compliance
Group 1
Class B
Not applicable
Not applicable
151
Specifications
5 m, Shield cable
Electromagnetic environment – guidance
The HMI-3000MT uses RF energy only for its internal
function. Therefore, its RF emissions are very low
and are not likely to cause any interference in nearby
electronic equipment.
The HMI-3000MT is suitable for use in all
establishments, including domestic establishments
and those directly connected to the public low-voltage
power supply network that supplies buildings used for
domestic purposes.