Adjustments / Programs
Connect the service PC to the serial interface on board D1.
Start the service program.
Select the Adjustment, Calibrations menu.
Answer the query "yes".
In the Combo box "Function Groups:" select "Camera Rotation" .
In the window "Available Functions:" select "1. Init (Disable Limits)" and click on
"Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit", click "No".
Attach the center cross, or other fluoroscopic image positioning element to the image
intensifier.
Start fluoro and rotate the camera to the +180° position
(With the camera rotation key clockwise).
The image contents should rotate counterclockwise.
Stop fluoro, in the window "Available Functions:", select "2. +180 Degree Position" and
click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit", click "No".
Start fluoro and set camera rotation to the 0° position.
The image (left/right / up/down) corresponds to Fig. 4
Stop fluoro, in the window "Available Functions:", select "3. 0 Degree Position" and click
on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit", click "No".
Start fluoro and rotate the camera to the -180° position (with the counterclockwise
camera rotation key). The image contents will rotate clockwise.
Stop fluoro, in the window "Available Functions:", select "4. -180 Degree Position" and
click on "Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit", click "No".
In the window "Available Functions:", select "5. Init Off & Store Positions" and click on
"Execute".
Close the status window by clicking on the OK button.
For the query "Do you want to reset the unit", click "Yes".
Return to the service program main menu with the ESC key.
Select the Data, Backup... menu.
Select Parameter in the Backup window.
Click on "Backup".
In "Remarks for Parameter", enter "camera rotation set".