9 Final tests

Remove the measurement devices.
PMF Operating problems
Were there any operating problems during the checks?
PMP Covers
Attach all removed covers. Check that all covers are complete and firmly seated, including the
cable ducts.
SIE
Protective conductor test
Perform the protective conductor test.
PMP Cleaning / Damaged paint
Clean the unit using the materials recommended in the Instructions for use and repair any
damaged paint.
System ON.
QSQ Final test exposure
Take one final test exposure.
MAMMOMAT Balance
Maintenance Instructions
SPB7-115.831.01
Page 21 of 22
Rev.01
05.04
9 Final tests