8 Miscellaneous

SIE
Testing the radiation field / light field
Proceed with the test according to the Installation & Start-Up instructions
SPB7-115.814.01... chapter "Checking the radiation field limitation".
PMF Automatic filter selection
If it exists select it from the control panel and check that audible sound during movement is
normal.
PMF RX signal lamp
Does the radiation-ON indicator light up?
PMF AEC Detector test
Check the Sensibility test [r] parameter as following:
During the test, the system should be configured as shown in the figure below.
Install Grid table 18x24
Install 18x24 collimation plate
Select LARGE focus
Remove compression plate
Remove cassette
Select ZERO POINTS technique
MAMMOMAT Balance
Maintenance Instructions
S I E M E N S
SPB7-115.831.01
Page 18 of 22
Rev.01
05.04
8 Miscellaneous