Capacitor formation (if necessary)

Remove the top cover
Switch "1" of S1 on D1 to "ON".
Select test programs by pressing "kV+" button until "P06" is displayed.
Press lamp button to start the capacitor formation.
Approximately every 6 minute the voltage values is increased. The charging sequence is
increased as follows:
At the 353 V level the charging will continue for another 48 minutes to decrease and stabi-
lize the capacitors internal leakage current.
Duration: Approximately 100 minutes.
Switch "1" of S1 on D1 to "OFF"
MOBILETT Plus HP
This will display the program version:
kVmAs
vErX.X = Version X.X (depends on software version installed).
The test program starts with displaying:
kVmAs
CAPFor = Capacitor formation program
After a short while, starts the program to display the actual voltage in the
capacitor bank:
kVmAs
CAP150 = 150 Volt.
150 - 250 - 299 - 325 - 337 - 345 - 348 - 351 - 353
If the formation is successfully finished the display will show:
kVmAs
CAP150 = 150 Volt.