Installation and setting

Function check

If a maIfunction occurs a message is indicated at the display, see "Service Instructions,
section Display messages".

Functional test cable winder

Disconnect the power plug
Pull the left brake handle (for the cable winder) upwards.
Pull it down and up until cable is completely rolled up.
Pull out the cable and connect the power plug.

Functional tests with motor drive

?
Test the forward/ reverse and fast (1 m/s) / slow (0.5 m/s) controls.
NOTICE

Speed Adjustment (if necessary)

Remove the plastic plug of tilt bar below the front cover (for details see service instruc-
tions).
Use a screwdriver and adjust the speed by potentiometer on D101:
- turn clockwise (CW) for slower speed.
- turn counterclockwise (CCW) for faster speed.
NOTICE
Siemens-Elema AB
Solna, Sweden
For adjustment, see chapter "Speed Adjustment".
This will effect the slow/fast speed simultaneously.