Installation

Preparatory work

Fig. 1 Removal of MOBILETT covers
1.
System OFF.
2.
Disconnect the mains cable.
3.
Remove the system upper cover (pos. 1 / Fig. 1). With MOBILETT Plus HP,
disconnect K106.
4.
Switch S10 to discharge position.
5.
Wait 5 minutes to allow the capacitor banks to discharge. Then remove the side
covers (pos. 2 / Fig. 1), the column cover (pos. 2 / Fig. 2) and the cover with the
cassette compartment (pos. 3 / Fig. 1).
6.
For MOBILETT Plus HP only: Remove the protective cover in front of the
batteries. Disconnect one of the blue battery connector pairs.