Fig. 6
Wiring the hand-held control unit
•
Guide the connector of the hand-held control unit from the front between the left
attachment rail and the power chain (as seen from the front) through to the rear.
•
Attach the connector of the hand-held control unit to the 9-pin SUB-D "HAND CTRL"
socket and secure it with screws.
•
Run the cable from the hand-held control unit to the cable holder (1/Fig. 6) and secure it
with a cable tie.
Wiring the printer
•
Insert the connector from the power cable in the power socket of the printer (2/Fig. 6).
•
The other end of the power cable is already connected to the ON/OFF assembly,
terminal block X2.1, X2.2 and the protective conductor terminal.
•
Insert the control cable in the printer socket and lock mechanically (3/Fig. 6).
•
Insert the power supply for the fans in the socket (4/Fig. 6).
SIREMOBIL Compact