P10 - Fast adaptation

This program is used after replacing the single tank or if the memory loses the pro-
grammed values. The user must make a number of exposures.
The generator calculates and stores the optimized filament values for each kV value from
these exposures.
After activating the program, the user must press the exposure button every time the
ready light indicator lights. The program starts at a low kV value and when the actual
value is stable, it increases to the next kV value automatically.
Example:
During the adaptation, some messages may appear, most of these can be ignored since
the generator has incorrect filament values. However, check every message against the
message list, and confirm them by pressing the "kV +" button. When the generator is
ready with an adaptation on a kV value, it continues to the next value. When the adapta-
tion is finished, the program will stop and the display will show:

P11 - Backup battery test

(Wiring diagram X037E, page 110)
After activating the program, it will display the battery voltage:
When the battery voltage is <2.8 V, the battery needs to be replaced. Refer to the section
"Replacement of backup battery".After 5 seconds, the program will be aborted. If the bat-
tery voltage is too low, there will be a message displayed with a message beep:
If the battery voltage is OK, the display will show:
MOBILETT Plus/ Plus E
kV
mAs
40
Ad Pr = Generator in adaptation mode
kV
mAs
P10 PASS = Adaptation program completed
kV
mAs
bAt 3.2 = 3.2 Volt
kV
mAs
P11 CAL 31
kV
mAs
P 11 PASS = Battery OK
SPR8-215.898.02
Rev. 01
01.02
Service programs
Page 6 of 8
Siemens-Elema AB
SPS-UD
5
5
Solna, Sweden