GE MobileLink CB-1000 Installation And Troubleshooting Manual page 10

Wireless communications for ecg
Table of Contents

Advertisement

6
Installation Procedures
5. Select the Network tab on the left side.
Enter the IP Address of the MAC 5000 client bridge and the
associated Netmask for this network segment.
127(
The Gateway address is only needed if the MAC 5000 will be on a
different subnet than the MUSE file server.
6. Select the Serial tab on the left side and then select the UART tab
from the top row of tabs.
7. Select/verify the following in the UART tab:
Baud
Data bits
Stop bits
Parity
MobileLink
2002783-060
115200
8 bits
1bit
none
03A
04A
Revision D