Carefully place board D1 on the 6 screws and secure it with washers and nuts (Fig.16).
Connect the plugs to the corresponding connections on board D1.
Secure the grounding strap to board D1, to the interface plate, to the heatsink and to the
base plate of the line filter.
When tightening the screws on board D1, be careful that the grounding strap does not
come in contact with adjacent components.
Secure all shielding of the previously connected cables to ground with cable clamps.
Secure the C-arm cable (1/Fig.16) and plug in the connectors.
Pull the Diamentor measurement cable through toward the left behind the lifting column.
Reattach the bracket (2/Fig.17) to the lifting column and tighten the screws.
Connect the footswitch cable to board D3 and secure it with the strain relief (Fig.18).