7.3.2 Changing the Density Adjustment Area

The density adjustment area (→ Setup Guide) can be changed manually in gray scale
adjustment mode (→ page 70) if [AUTO ADJUST.].
1
2
3
Annotations
When you change the image trimming range, annotations (fixed or free) may be out of
the image.
Exit the screen.
Touch [OK].
Select [ADJUST ROI].
Touch [ADJUST TRIM]/[ADJUST ROI] and select [ADJUST ROI].
Set the density adjustment area.
Touch two points on the image. The area you have selected will be indicated by a reddish
purple frame.
The reference point is changed, and the density of the entire image changes.
Exit the screen.
Touch [OK].
Changing the gray scale
When [AUTO ADJUST.] is touched to change to [ADJUST REX], [ADJUST TRIM]/
[ADJUST ROI] is fixed at [ADJUST TRIM] and the density adjustment area cannot
be changed.
7. Processing the Image (QA Mode Screen)
Density Adjustment Area
– 69 –