Battery - GE Kelman Minitrans Operator's Manual

Transformer oil dissolved gas and moisture monitor
Hide thumbs Also See for Kelman Minitrans:
Table of Contents

Advertisement

Relay 2 notifies the user when a user-configurable caution and/or alarm threshold is
reached.
The red LED indicates a caution and/or alarm state. Flashing denotes a caution state
whereas continuous illumination denotes an alarm state. If after the next
measurement run, no alarm or caution states exist, the red LED switches off. Use the
TransConnect software to set alarm and caution limits for many different
combinations of results. (See the TransConnect user guide for more information).
5.3

Battery

The product uses a non-rechargeable lithium coin cell battery (Panasonic CR2450 3 V
620 mAh) as shown in Figure 5-3. If the battery needs to be replaced, data from the
product must be backed up. Failure to do so may result in historical data loss.
Figure 5-3: Coin cell battery
The following steps describe how to change the battery:
1.
Back up the product data – contact your GE representative.
2.
Open the inner door to locate the battery on the system board (as shown in Figure
5-3).
3.
Slide the battery out of its housing.
4.
Replace with a new Panasonic CR2450 3 V 620 mAh coin cell.
5.
Close the inner door.
There is a danger of a new battery exploding if installed incorrectly.
Dispose of the used battery in accordance with local regulations — not in a
fire or with household waste. Contact your local waste disposal agency for the
address of the nearest battery deposit site. Perchlorate material — special
handling may apply.
See:
MA-006- Minitrans Operator Guide - Rev 2.2
www.dtsc.ca.gov/hazardouswaste/perchlorate/
5-Mar-20
Page 11 of 26

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents