Table 8-14
2.
Disconnect the cables to the Monitor.
NOTE: If you are replacing any cables (1),
you will also need to remove two additional
tie-wraps (2) at the bracket.
Remove the clamps (3).
NOTE: If the device is equipped with the
WLAN option, remove the WLAN dongle
and transfer it to the new monitor.
To remove the dongle, slightly lift up the
end to disengage the Velcro, then pull the
dongle out of the USB port.
3.
Tilt the monitor back to vertical position.
Lock the monitor back into place vertically.
Loosen the four screws by turning each
screw between one half and one turn
counter-clockwise. You don't need to
remove the screws.
Lift the LCD Monitor assembly upwards
until you can lift it away from the Monitor
Bracket.
Place the LCD Monitor on a clean and safe
surface.
5
EV
LCD Monitor removal - R3.x and earlier
Steps
Chapter 8 Replacement procedures
LOGIQ E9 Service Manual
Corresponding Graphic
LCD Monitor Cables and Clamps
WLAN Dongle removal
LCD Monitor fixing screws