Step
5
6
7
Note:
ÄKTA go Operating Instructions 29360951 AC
Action
Specify the details of the result file from the performance test. Click Start to
start the performance test.
Note:
The result file is separate from the test report. The report will be generated
even if No result is selected.
Follow any instructions that are shown on the screen.
Check whether the test was passed or failed in the System Performance
Report, which is located at C:\Program Files (x86)\GE Healthcare\UNI-
CORN\UNICORN x.x\Temp, where x.x is the UNICORN version number. The
test result is stated at the top of the report. If the test was failed refer to
ÄKTA go User Manual for possible causes.
The progress of the performance test is shown in the Chromatogram pane
in the System Control module.
5 Prepare the system for a run
5.3 Performance tests
73