i. Remove Steering Column (incl. laying cables)
Before the steering column and the cables (spiral hose + cable connecting control terminal to PCB
Interface) can be removed, PCB Interface, the pneumatic system of the BP module and the power
supply must be removed.
•
Remove the control terminal.
•
Disconnect the cable connecting the control terminal
to PCB Interface Messkopf X150.
•
Disconnect NIBP tubing.
•
Open cable ties of connection cable and spiral hose
(see illustrations).
•
Pull connection cable and spiral hose towards the
steering column.
•
Loosen clamping lever (eBike comfort) securing the
handlebar or
loosen fixation of the steering column (eBike basic).
Caution: ultra-tight screw lock.
In each case take care not to drop the clamping plate.
•
Lift off the steering column.
Note:
When installing the steering column with the spiral hose and
connection cable, cable and hose must be slack and secured
with cable ties (see illustrations).
j. Remove Drive Unit
•
Disconnect connection cable of PCB RPM X200 from PCB Interface.
•
Unscrew the 6 screws (M8x20) of the drive unit
and lower the drive unit.
k. Remove PCB RPM
•
Disconnect the PCB RPM cable J3.
•
Remove connection J5 to strain gauge.
•
Unscrew the two screws of PCB RPM.
•
Disconnect the two cables (red + blue) to the eddy-current brake.
2018112-003 Rev J
B: Mechanical Design
eBike, eBike L, eBike EL
- 19 -