Chapter 4 Command Reference
:INITiate Subsystem
:INITiate Subsystem
This subsystem controls the initiation of a measurement.
:INITiate:CONTinuous . . . <Boolean>
Sets or queries the state of continuously initiated measurements.
When CONTinuous is set to OFF, no measurements are made until
CONTinuous is set to ON or :INITiate[:IMMediate] is received. Once
CONTinuous is set to ON, a new measurement is initiated. On the
completion of each measurement, with CONTinuous ON, another
measurement immediately commences.
Single ASCII-encoded byte, 0 or 1.
A value of 0 indicates OFF; a value of 1 indicates ON.
*RST: OFF
The commencement of the first measurement due to setting
:INITiate:CONTinuous to ON sets the Pending Operation Flag to
true. The Pending Operation Flag is set false by aborting a
measurement, or by the completion of the last measurement after
:INITiate:CONTinuous is set OFF.
With the measurements being made continuously, the :ABORt
command aborts the current measurement in progress, however, the
value of :INITiate:CONTinuous is unaffected. If CONTinuous was set
to ON prior to receiving :ABORt, it remains ON and a new
measurement begins.
When a single measurement is in progress (:INIT:CONT is OFF):
Error -213 (Init ignored) is generated and the state of INIT:CONT
is unaffected by :INIT:CONT ON.
Error -210 (Trigger error) is generated by INIT:CONT OFF.
The Counter powers up with :INIT:CONT set to ON, but *RST sets
:INIT:CONT to OFF.
4-8
Programming Guide