Chapter 4 Command Reference
Group Execute Trigger (GET)
Group Execute Trigger
(GET)
GET
COMMAND
The full capability of the Group Execute Trigger IEEE 488.1 interface
function is implemented in the Counter. This function permits the
Counter to have its operation initiated over the Bus. In response to the
IEEE 488.1 Group Execute Trigger (GET) remote interface message
(while the Counter is addressed to listen), the Counter performs the action
defined by the *DDT command (see page 4-46).