Check speed indication on LCD and additional speed indication for test subject.
Check noise level of idling drive unit (e.g. for grinding noises or noise from the bearing)
Expected result: Speed displayed on both displays and no noticeable noise
•
Units with NIBP module:
- Check cuff tubing and tubing connections, microfon connectors
- Check blood pressure measurement on test subject at rest
- Check blood pressure readings on LCD
Expected result:
•
Start manual exercise test, change load (e.g. start 50 W, increase 25 W)
- Check load level and load indication
Expected result:
•
If ergometer is remote controlled from ECG recorder or PC-ECG system:
- Check connecting cable
- Check load control via ECG recorder or PC ECG system
- Check remote start if applicable
Expected result:
2018112-069 Rev. F
BP values are displayed and in proper range
Load changes as expected
Ergometer can be controlled by external unit
eBike II, eBike II L, eBike II EL
Pass/Fail
Pass/Fail
Pass/Fail
Pass/Fail