Set the calibrator unit to 110 rpm.
•
Wait approx. 20 seconds, until the load display is
stable
•
By using the arrow keys, input the actual value of the
calibrator (e.g. 55,0 Watt) at the ergometer.
•
Store the value with SET.
•
Set the ergometer to 400 Watt (via PC interface)
•
Wait approx. 20 seconds, until the load display is
stable
•
By using the arrow keys, input the actual value of the
calibrator (e.g. 412,5 Watt) at the ergometer.
•
Store the value with SET.
The successfull completion of the dynamic load calibration
is confirmed.
•
Stop the calibration unit.
- 70 -
Dynamic Load Cal.
Set at 110 rpm / 50 Watt
Rotation:
DMS-Volt:
LRE Load:
Dyn. Load
d
ynaMic load calibration
Dynamic Load Cal.
Set at 110 rpm / 400 Watt
Rotation:
DMS-Volt:
LRE Load:
Dyn. Load
d
ynaMic load calibration
Dynamic Load Cal.
Rotation:
DMS-Volt:
LRE Load:
d
ynaMic load calibration
eBike II, eBike II L, eBike II EL
50
327
50
55,0
SET
- 50 w
att at
50
327
400
412,5
SET
- 400 w
att at
110
1327
400
Calibration successfull
-
finiShed
2018112-069 Rev. F
mV
Watt
Watt
110
rpM
mV
Watt
Watt
110
rpM
mV
Watt