Analog Calibration - GE eBike III basic Series Service Manual

Electric bike with and without nibp
Table of Contents

Advertisement

nIBp t
/ nIBp C
est
alIBratIon
This menu is enabled, if the ergometer is equipped with an
automatic NIBP module and this module is activated.
Test and calibration of the NIBP module is described in
session „Calibration of Blood pressure unit" on page 75..
a
C
nalog
alIBratIon
This function is used to allign the analog remote signals
(input and output) to the corresponding ECG devices.
The eBike III ergometers have to be equipped with the
optional „COM Module (PN 201118-340)". This module is
connected with a small cable to the 9-pin DSUB connector
and fixed to the chassis with a plastic velcro tape.
In addition, a jummper connector has to be installed on
the LIU board.
The COM module handles one analog input and five analog
output signals from/to external devices.
Select the analog signal to be aligned and confirm
by pressing the NIBP key.
- 68 -
To align the analog load signals, the corresponding ECG
recorder should be connected as a reference voltage
source
Disconnect the RS232 or USB cable, if connected
The alignment of the analog signals can be controlled
after exiting the service menu (e g by starting a remote
controlled stress test)
S
ignal aSSignment of analog connector
Pin 1 :
Pin 2:
Pin 3:
Pin 4:
Pin 5:
Pin 6:
Pin 7:
Pin 8:
Housing:

Analog Calibration

Input
Output
Output
Output
Output
EXIT
a
lignment of analog input
eBike III basic / comfort
Note
(com
INPUT Load
reserved - do not connect
reserved - do not connect
OUTPUT BP heartrate
OUTPUT Load
OUTPUT Diastole
OUTPUT rpm
OUTPUT Systole
Signal GND
1
Load
1
Load
2
Heart.
3
Syst.
4
Diast.
/
output SignalS
2018111-336 Rev. B
)