Electrical Safety - GE eBike III basic Series Service Manual

Electric bike with and without nibp
Table of Contents

Advertisement

e
s
leCtrICal
afety
Perform electrical safety checks when indicated - current leakage test results meet requirements.
All indicated electrical safety checks require a pass/fail indication for steps performed.
Record the measurement values in your debrief.
Step
Earth Leakage Current
1.
2.
Neutral open, Forward Polarity
3.
Neutral open, Reverse Polarity
4.
Enclosure Leakage Current
1.
2.
Neutral open, Forward Polarity
3.
Ground open, Forward Polarity
4.
Ground open, Reverse Polarity
5.
Neutral open, Reverse Polarity
6.
Patient Leakage Current to Ground (3)
1.
2.
Neutral open, Forward Polarity
3.
Ground open, Forward Polarity
4.
Ground open, Reverse Polarity
5.
Neutral open, Reverse Polarity
6.
Ground Continuity
1.
AC mains power cord ground prong to
exposed metal surface (ground lug)
(1) NC =
Normal Condition
SFC =
Single Fault Condition
N/A =
Not Applicable
(2) UUT =
Unit Under Test
(3) Test applies for eBike III with blood pressure unit only - measuring point is the microphone connector
- 10 -
Condition
(1)
Forward Polarity
N/A
N/A
N/A
Reverse Polarity
N/A
Forward Polarity
NC
SFC
N/A
N/A
SFC
Reverse Polarity
NC
Forward Polarity
NC
SFC
N/A
N/A
SFC
Reverse Polarity
NC
N/A
eBike III basic / comfort
UUT - ON
Result
(2)
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
__________ μA
Pass/Fail
Leakage
Current limits
100 μA
500 μA
500 μA
100 μA
100 μA
5000 μA
5000 μA
100 μA
Resistance
2018111-336 Rev. B