EX Safety Instructions Mass Flowmeter SI/FCM2000/FM/CSA-EN CoriolisMaster FCM2000 cFMus Div. 1, Div. 2 Safety information for electrical equipment for potentially explosive areas in accordance with Factory Mutual Standards and Canadian Standards Association Mass Flowmeter...
The permissible temperature data and limit values Safety information for electrical installation Never attempt electrical connection unless the power supply is switched off. Ground the flowmeter sensor and transmitter housing. Make sure there is no risk of explosion. CoriolisMaster FCM2000 SI/FCM2000/FM/CSA-EN...
Failure to observe this safety information may result in damage to or destruction of the product and/or other system components. IMPORTANT (NOTE) This symbol indicates operator tips, particularly useful information, or important information about the product or its further uses. It does not indicate a dangerous or damaging situation. SI/FCM2000/FM/CSA-EN CoriolisMaster FCM2000...
Control Install Dwg No. MDM-10-A0291 Size... Tamb. = -40°C...+60°C Type 4X, IP 6X G00354-01 Fig. 1 1 Order no. 5 Supply power and maximum power 2 FM approval consumption 3 Nominal diameter 6 Ingress protection 4 Ambient temperature CoriolisMaster FCM2000 SI/FCM2000/FM/CSA-EN...
− Class I Div. 1 ME24/25 T, X ... ME24/25 O, V, P, W MS26 C, Y, D, Q 2. Remote mount design (small nominal diameters) Flowmeter sensor − Class I Div. 1 G00899 Fig. 2: FCM2000 overview SI/FCM2000/FM/CSA-EN CoriolisMaster FCM2000...
However, it is also possible to supply devices certified in accordance with ATEX or IECEx that feature 1/2" NPT threads without glands. In such cases, the user is responsible for ensuring that the cable piping/glands are installed in accordance with the relevant national standards (e.g., NEC, CEC, ATEX 137, IEC 60079-14, etc.). CoriolisMaster FCM2000 SI/FCM2000/FM/CSA-EN...
3. Slide the green/yellow protective tubing over the strand, leaving 10 mm (0.39 inch) protruding from the end (if necessary, shorten the protective tubing). 4. Press on the ring-type terminal (2) supplied. 5. Connect to the grounding connection (3). G00900 Fig. 3 SI/FCM2000/FM/CSA-EN CoriolisMaster FCM2000...
Groups E, F and G shall be limited to a maximum surface temperature of 165°C (329 °F). Seal all conduits within 18 inches (457.20 mm). The flow meters from ABB are designed for the world wide industrial market and are suitable to measure among others flammable and combustible process fluids and are installed within a process pipe.
Type: MS2 in Class I Div. 1 or Class I Div. 2 Ambient temperature -20 ... 50 °C (-4 ... 122 °F) Temperature class Maximum permissible fluid temperature 180 °C (356 °F) 180 °C (356 °F) 180 °C (356 °F) 125 °C (257 °F) 80 °C (176 °F) SI/FCM2000/FM/CSA-EN CoriolisMaster FCM2000...
Digital input 0,17 Terminal 81/82 Pulse output 0,17 Terminal 51/52 All inputs and outputs are electrically isolated from each other and from the supply power. Only current outputs 1 and 2 are not electrically isolated from one another. CoriolisMaster FCM2000 SI/FCM2000/FM/CSA-EN...
Page 13
The housing for the transmitter and flowmeter sensor must be connected to the potential equalization PA. The operator must ensure that when connecting the protective conductor (PE) no potential differences can occur between protective conductor and potential equalization (PA). SI/FCM2000/FM/CSA-EN CoriolisMaster FCM2000...
The housing for the transmitter and flowmeter sensor must be connected to the potential equalization PA. The operator must ensure that when connecting the protective conductor (PE) no potential differences can occur between protective conductor and potential equalization (PA). CoriolisMaster FCM2000 SI/FCM2000/FM/CSA-EN...
• The flowmeter sensor must be connected to the transmitter by means of the signal cable supplied by ABB Automation Products; no other cable may be used for this purpose. • To ensure safe operation it is imperative that the information in this manual is observed.
Whatever the application, devices with transmitter type ME23/28 must be connected and installed in the same way as in Div. 1. The electrical connection between the flowmeter sensor and transmitter must be established in accordance with the information provided in the associated operating instructions (OI/FCM2000). CoriolisMaster FCM2000 SI/FCM2000/FM/CSA-EN...
The electronic components of the printed circuit board can be damaged by static electricity (observe ESD guidelines). Make sure that the static electricity in your body is discharged before touching electronic components. IMPORTANT (NOTE) Users must also observe the information in the operating and commissioning instruction for the device. SI/FCM2000/FM/CSA-EN CoriolisMaster FCM2000...
Designation relating to intended use in potentially explosive atmospheres in protection compliance with: ATEX Directive (marking in addition to CE marking) IEC standards Approvals for Canada and United States IMPORTANT (NOTE) All documentation, declarations of conformity and certificates are available in ABB's download area. www.abb.com/flow CoriolisMaster FCM2000 SI/FCM2000/FM/CSA-EN...