Section 2: Installation

Overview

This section provides detailed installation and setup instructions for the Hematek Slide
Stainer. The installation steps must be followed correctly to ensure proper installation,
operation, and service. Read this Operator's Guide carefully before attempting to operate
the instrument. Follow all instructions carefully.
The Hematek Slide Stainer is a precision instrument and must be handled accordingly.
Rough handling or dropping of the instrument will disturb or damage internal
components. Always handle the instrument with care.

Environmental Factors

As with all sensitive electronic instruments, prolonged exposure to excessive humidity
and temperature should be avoided. Temperature should be held relatively constant to
obtain the highest degree of operating stability. The ambient temperature range for
operating the instrument is 18ºC to 30ºC (64ºF to 86ºF). The ambient operating humidity
range is 20% to 85% relative humidity.
Place the instrument in a well-ventilated area, avoiding exposure to corrosive vapors or
temperature extremes. Be sure it is near a power source that meets the electrical
requirements (voltage) specified on the rating label located on the rear of the instrument.
Avoid proximity to open windows, sinks, ovens, hot plates, open burners, radiators, and
dry ice baths. The instrument should not be used in an explosive atmosphere.