4.1 Startup Sequence

By default, the Giraffe Incubator starts up in user mode. The next section describes how to start up the
Incubator in service mode. The following table describes the Giraffe Incubator startup sequence for each mode:
User Mode
1. The alarm light turns on and remains on.
2. All LEDs and segments light up and the EL
screen is yellow.
3. All LEDs turn off. The EL screen displays the
Giraffe logo.
4. The alarm light turns off. Software revisions
appear on the EL screen with OKs.
5. The user screen appears.
© 2001 by Datex-Ohmeda, Inc.. All rights reserved.

Chapter 4: Troubleshooting

Service Mode
1. The alarm light turns on and remains on.
2. All LEDs and segments light up and the EL screen is
yellow.
3. All LEDs turn off. The EL screen displays the Giraffe
logo.
4. The alarm light turns off. Software revisions appear
on the EL screen without OKs.
5. The service screen appears.