Section 6 Product Overview

Purge

The presence of moisture is no longer an issue with the MB3000 Series FTIR
spectrometer since it is designed with non hygroscopic materials. However many
compressors tend to let some oil mist enter the air that is being compressed. Oil mist
is not removed by moisture-carbon dioxide scrubbing filters. Therefore, use dry oil-
free air or nitrogen only to purge the system.

Power Inlet

Power cords are provided in accordance with the destination country.