Configuring Anesthesia Machine to Mini Field Agent
18
For example, if the UTC timestamp is Wed June 7 9:13:54
2017, type date -s "2017-06-07 09:13:54".
•
hwclock --systohc
This command updates the hardware clock to local system
time.
2.
To verify that the correct timestamp and timezone is set and to
check if the hardware clock is in sync with the local time, type
the following and then press
•
date
•
hwclock --show
Perform the following procedure to configure the anesthesia machine
to communicate with the SBX Network Client. See Super User mode
in the Anesthesia Machine User's Reference Manual for further
details.
Note
This configuration has to be performed on each anesthesia machine
that communicates to the Mini Field Agent.
1.
Enter Super User mode on the anesthesia machine.
2.
System Config
Select
the set time Zone.
Note
Verify that the Time Zone set is based on the region where
anesthesia machine is installed. The Time Zone set on the
anesthesia machine for New York is shown for reference in the
below screenshot.
Enter
after each entry.
Time Zone
and then select
2094532-001 D 02 2018
to view