HP 5890 SERIES II Service Manual page 214

Gas chromatograph
Hide thumbs Also See for 5890 SERIES II:
Table of Contents

Advertisement

36. Install the column and any other associated hardware removed in step 2 of this procedure.
37. Install the collector mount and collector assem-
bly as a unit.
38. Secure the collector mount to the thermal strap
using three screws.
39. Connect the ignitor wire lead connector at the
mating connection adjacent to the ignitor.
40. Slide the rear cover on to the instrument.
41. Secure the rear cover to the instrument by
installing and tightening four screws.
42. Install the right side panel and secure
using four screws.
43. Install the electronics carrier top cover.
44. Install the hinged top cover and secure using a screw and washer.
45. Connect the ground strap to the hinged top cover using a screw.
46. Restore all gas supplies.
47. Restore power to the instrument.
Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com
COLLECTOR
NUT
COLLECTOR
SPANNER
NUT
COLLECTOR
MOUNT
THERMAL
STRAP
SVC 6- 23
DETECTOR PCB
INTERCONNECT
IGNITOR
CLAMPS

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents