28. Install the liner in the shell weldment.
29. Install a cap or plug on the end of the inlet (inside the column oven).
30. Restore the supply gas pressure.
31. Check for leaks at all of the newly mated fittings.
32. Turn off the supply gas.
33. Remove the cap/plug from the end of the inlet.
34. Install the column and associated hardware removed in step 2.
35. Install the right side panel and secure using two screws.
36. Install the electronics carrier top cover.
37. Slide the rear cover on to the instrument.
38. Secure the rear cover to the instrument by installing and tightening four screws.
39. Install the injection port cover.
40. Restore power to the HP 5890 Series II.

SVC 6- 15

Artisan Scientific - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisan-scientific.com